Hamilton Medical AG

0 FDA warning letters since 2023 and 2 recalls in the last 12 months.

Get an email the day FDA posts anything new about Hamilton Medical AG.

Recalls

ReportedClassFirmProduct and reason
Jul 8 Class I Hamilton Medical AG
Device
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement
Jul 8 Class II Hamilton Medical AG
Device
IntelliCuff, Portable automatic cuff pressure controller REF: 951001
Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.