Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Jul 8 | Class I | Hamilton Medical AG Device |
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators. Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement |
| Jul 8 | Class II | Hamilton Medical AG Device |
IntelliCuff, Portable automatic cuff pressure controller REF: 951001 Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user. |